■ Demo— not a validated GxP system — synthetic data

Affinis

AFN-000520-01 · EGFR L858R/T790M

Rejected

CRV-2026-00403-025 · v1 · auto-fitPlates PL-2026-0403, PL-2026-0404 (n = 2 replicates)4PL · Levenberg–Marquardt · unweighted, log₁₀[C]

Curve 7 / 10 in batch

Response vs concentration

  • Rep. 1 · PL-0403
  • Rep. 2 · PL-0404
  • Masked
  • IC50 95 % CI
Response against concentration, EGFR L858R/T790M−10−9−8−7−6−50255075100log₁₀ [compound] (M)% inhibitionM03 on PL-2026-0403M04 on PL-2026-0403M05 on PL-2026-0403M06 on PL-2026-0403M07 on PL-2026-0403M08 on PL-2026-0403M09 on PL-2026-0403M10 on PL-2026-0403M11 on PL-2026-0403M12 on PL-2026-0403M03 on PL-2026-0404M04 on PL-2026-0404M05 on PL-2026-0404M06 on PL-2026-0404M07 on PL-2026-0404M08 on PL-2026-0404M09 on PL-2026-0404M10 on PL-2026-0404M11 on PL-2026-0404M12 on PL-2026-0404

Concentrations: 10-point, 1:3, 0.508 nM – 10.0 µMPoints used 20 / 20

Fit parameters — v1 · auto-fit

ParameterEstimate95 % CIConstraint
Top100.0 %[97.1, 103.0] %Free
Bottom6.6 %[4.7, 8.4] %Free
IC50193 nM[171, 219] nMFree
pIC506.71± 0.03 (SE)Derived
Hill slope (nH)0.99[0.88, 1.11]Free

R² 0.998RMSE 1.7 %df 16Fit engine 1.2.0 · 20 / 20 points

Review decision

IC50

193 nM

95 % CI 171–219 nM · pIC50 6.71

Rejected: Curve does not plateau — retest over a wider range

History

append-only
  1. v1 auto-fit createdsystem · 2026-11-04 17:05:40 UTCFit engine 1.2.0 · 20 / 20 points · no flags
  2. Rejectedmlaurent · 2026-11-04 18:10:00 UTC“Curve does not plateau — retest over a wider range” · v1 · auto-fit