■ Demo— not a validated GxP system — synthetic data

Affinis

AFN-000836-01 · EGFR L858R/T790M

Rejected

CRV-2026-00407-001 · v1 · auto-fitPlates PL-2026-0407, PL-2026-0408 (n = 2 replicates)4PL · Levenberg–Marquardt · unweighted, log₁₀[C]

Curve 7 / 10 in batch

Response vs concentration

  • Rep. 1 · PL-0407
  • Rep. 2 · PL-0408
  • Masked
  • IC50 95 % CI
Response against concentration, EGFR L858R/T790M−10−9−8−7−6−50255075100log₁₀ [compound] (M)% inhibitionA03 on PL-2026-0407A04 on PL-2026-0407A05 on PL-2026-0407A06 on PL-2026-0407A07 on PL-2026-0407A08 on PL-2026-0407A09 on PL-2026-0407A10 on PL-2026-0407A11 on PL-2026-0407A12 on PL-2026-0407A03 on PL-2026-0408A04 on PL-2026-0408A05 on PL-2026-0408A06 on PL-2026-0408A07 on PL-2026-0408A08 on PL-2026-0408A09 on PL-2026-0408A10 on PL-2026-0408A11 on PL-2026-0408A12 on PL-2026-0408

Concentrations: 10-point, 1:3, 0.508 nM – 10.0 µMPoints used 20 / 20

Fit parameters — v1 · auto-fit

ParameterEstimate95 % CIConstraint
Top99.1 %[97.3, 100.9] %Free
Bottom−2.8 %[−4.9, −0.7] %Free
IC5050.1 nM[45.8, 54.7] nMFree
pIC507.30± 0.02 (SE)Derived
Hill slope (nH)1.24[1.11, 1.36]Free

R² 0.998RMSE 1.7 %df 16Fit engine 1.2.0 · 20 / 20 points

Review decision

IC50

50.1 nM

95 % CI 45.8–54.7 nM · pIC50 7.30

Rejected: Curve does not plateau — retest over a wider range

History

append-only
  1. v1 auto-fit createdsystem · 2026-11-04 17:05:40 UTCFit engine 1.2.0 · 20 / 20 points · no flags
  2. Rejectedmlaurent · 2026-11-04 18:10:00 UTC“Curve does not plateau — retest over a wider range” · v1 · auto-fit