■ Demo— not a validated GxP system — synthetic data

Affinis

AFN-001446-01 · EGFR WT

Rejected

CRV-2026-00450-002 · v1 · auto-fitPlates PL-2026-0450, PL-2026-0451 (n = 2 replicates)4PL · Levenberg–Marquardt · unweighted, log₁₀[C]

Curve 2 / 10 in batch

Response vs concentration

  • Rep. 1 · PL-0450
  • Rep. 2 · PL-0451
  • Masked
  • IC50 95 % CI
Response against concentration, EGFR WT−10−9−8−7−6−50255075100log₁₀ [compound] (M)% inhibitionA13 on PL-2026-0450A14 on PL-2026-0450A15 on PL-2026-0450A16 on PL-2026-0450A17 on PL-2026-0450A18 on PL-2026-0450A19 on PL-2026-0450A20 on PL-2026-0450A21 on PL-2026-0450A22 on PL-2026-0450A13 on PL-2026-0451A14 on PL-2026-0451A15 on PL-2026-0451A16 on PL-2026-0451A17 on PL-2026-0451A18 on PL-2026-0451A19 on PL-2026-0451A20 on PL-2026-0451A21 on PL-2026-0451A22 on PL-2026-0451

Concentrations: 10-point, 1:3, 0.508 nM – 10.0 µMPoints used 20 / 20

Fit parameters — v1 · auto-fit

ParameterEstimate95 % CIConstraint
Top99.2 %[92.5, 105.9] %Free
Bottom−3.6 %[−5.3, −1.9] %Free
IC501.18 µM[0.992, 1.39] µMFree
pIC505.93± 0.03 (SE)Derived
Hill slope (nH)1.19[1.01, 1.38]Free

R² 0.996RMSE 2.1 %df 16Fit engine 1.2.0 · 20 / 20 points

Review decision

IC50

1.18 µM

95 % CI 0.992–1.39 µM · pIC50 5.93

Rejected: Curve does not plateau — retest over a wider range

History

append-only
  1. v1 auto-fit createdsystem · 2026-11-04 17:05:40 UTCFit engine 1.2.0 · 20 / 20 points · no flags
  2. Rejectedmlaurent · 2026-11-04 18:10:00 UTC“Curve does not plateau — retest over a wider range” · v1 · auto-fit