■ Demo— not a validated GxP system — synthetic data

Affinis

AFN-004100-01 · EGFR WT

Rejected

CRV-2026-00472-018 · v1 · auto-fitPlates PL-2026-0472, PL-2026-0473 (n = 2 replicates)4PL · Levenberg–Marquardt · unweighted, log₁₀[C]

Curve 10 / 10 in batch

Response vs concentration

  • Rep. 1 · PL-0472
  • Rep. 2 · PL-0473
  • Masked
  • IC50 95 % CI
Response against concentration, EGFR WT−10−9−8−7−6−50255075100log₁₀ [compound] (M)% inhibitionI13 on PL-2026-0472I14 on PL-2026-0472I15 on PL-2026-0472I16 on PL-2026-0472I17 on PL-2026-0472I18 on PL-2026-0472I19 on PL-2026-0472I20 on PL-2026-0472I21 on PL-2026-0472I22 on PL-2026-0472I13 on PL-2026-0473I14 on PL-2026-0473I15 on PL-2026-0473I16 on PL-2026-0473I17 on PL-2026-0473I18 on PL-2026-0473I19 on PL-2026-0473I20 on PL-2026-0473I21 on PL-2026-0473I22 on PL-2026-0473

Concentrations: 10-point, 1:3, 0.508 nM – 10.0 µMPoints used 20 / 20

Fit parameters — v1 · auto-fit

ParameterEstimate95 % CIConstraint
Top100.7 %[100.0, 101.5] %Free
Bottom7.7 %[6.1, 9.4] %Free
IC5012.6 nM[11.9, 13.4] nMFree
pIC507.90± 0.01 (SE)Derived
Hill slope (nH)1.06[1.00, 1.12]Free

R² 0.999RMSE 0.8 %df 16Fit engine 1.2.0 · 20 / 20 points

Review decision

IC50

12.6 nM

95 % CI 11.9–13.4 nM · pIC50 7.90

Rejected: Curve does not plateau — retest over a wider range

History

append-only
  1. v1 auto-fit createdsystem · 2026-11-04 17:05:40 UTCFit engine 1.2.0 · 20 / 20 points · no flags
  2. Rejectedmlaurent · 2026-11-04 18:10:00 UTC“Curve does not plateau — retest over a wider range” · v1 · auto-fit