■ Demo— not a validated GxP system — synthetic data

Affinis

AFN-001174-01 · H1975 cells

Rejected

CRV-2026-00492-002 · v1 · auto-fitPlates PL-2026-0492 (n = 1 replicates)4PL · Levenberg–Marquardt · unweighted, log₁₀[C]

Curve 10 / 10 in batch

Response vs concentration

  • Rep. 1 · PL-0492
  • Masked
  • IC50 95 % CI
Response against concentration, H1975 cells−9−8−7−6−50255075100log₁₀ [compound] (M)% inhibitionB03 on PL-2026-0492B04 on PL-2026-0492B05 on PL-2026-0492B06 on PL-2026-0492B07 on PL-2026-0492B08 on PL-2026-0492B09 on PL-2026-0492B10 on PL-2026-0492

Concentrations: 8-point, 1:3, 4.57 nM – 10.0 µMPoints used 8 / 8

Fit parameters — v1 · auto-fit

ParameterEstimate95 % CIConstraint
Top97.6 %[87.1, 108.0] %Free
Bottom5.7 %[2.8, 8.5] %Free
IC501.42 µM[1.07, 1.88] µMFree
pIC505.85± 0.04 (SE)Derived
Hill slope (nH)1.14[0.86, 1.43]Free

R² 0.999RMSE 1.1 %df 4Fit engine 1.2.0 · 8 / 8 points

Review decision

IC50

1.42 µM

95 % CI 1.07–1.88 µM · pIC50 5.85

Rejected: Signal drift across the plate, replicates disagree

History

append-only
  1. v1 auto-fit createdsystem · 2026-11-04 17:05:40 UTCFit engine 1.2.0 · 8 / 8 points · no flags
  2. Rejectedmlaurent · 2026-11-04 18:10:00 UTC“Signal drift across the plate, replicates disagree” · v1 · auto-fit