■ Demo— not a validated GxP system — synthetic data

Affinis

AFN-002631-01 · H1975 cells

Rejected

CRV-2026-00515-008 · v1 · auto-fitPlates PL-2026-0515 (n = 1 replicates)4PL · Levenberg–Marquardt · unweighted, log₁₀[C]

Curve 10 / 10 in batch

Response vs concentration

  • Rep. 1 · PL-0515
  • Masked
  • IC50 95 % CI
Response against concentration, H1975 cells−9−8−7−6−50255075100log₁₀ [compound] (M)% inhibitionH03 on PL-2026-0515H04 on PL-2026-0515H05 on PL-2026-0515H06 on PL-2026-0515H07 on PL-2026-0515H08 on PL-2026-0515H09 on PL-2026-0515H10 on PL-2026-0515

Concentrations: 8-point, 1:3, 4.57 nM – 10.0 µMPoints used 8 / 8

Fit parameters — v1 · auto-fit

ParameterEstimate95 % CIConstraint
Top97.7 %[87.7, 107.8] %Free
Bottom0.8 %[−2.0, 3.7] %Free
IC501.47 µM[1.15, 1.88] µMFree
pIC505.83± 0.04 (SE)Derived
Hill slope (nH)1.27[0.96, 1.58]Free

R² 0.999RMSE 1.2 %df 4Fit engine 1.2.0 · 8 / 8 points

Review decision

IC50

1.47 µM

95 % CI 1.15–1.88 µM · pIC50 5.83

Rejected: Curve does not plateau — retest over a wider range

History

append-only
  1. v1 auto-fit createdsystem · 2026-11-04 17:05:40 UTCFit engine 1.2.0 · 8 / 8 points · no flags
  2. Rejectedmlaurent · 2026-11-04 18:10:00 UTC“Curve does not plateau — retest over a wider range” · v1 · auto-fit