■ Demo— not a validated GxP system — synthetic data

Affinis

AFN-003320-01 · H1975 cells

Approved · locked

CRV-2026-00527-003 · v1 · auto-fitPlates PL-2026-0527 (n = 1 replicates)4PL · Levenberg–Marquardt · unweighted, log₁₀[C]

Curve 1 / 10 in batch

Response vs concentration

  • Rep. 1 · PL-0527
  • Masked
  • IC50 95 % CI
Response against concentration, H1975 cells−9−8−7−6−50255075100log₁₀ [compound] (M)% inhibitionC03 on PL-2026-0527C04 on PL-2026-0527C05 on PL-2026-0527C06 on PL-2026-0527C07 on PL-2026-0527C08 on PL-2026-0527C09 on PL-2026-0527C10 on PL-2026-0527

Concentrations: 8-point, 1:3, 4.57 nM – 10.0 µMPoints used 8 / 8

Fit parameters — v1 · auto-fit

ParameterEstimate95 % CIConstraint
Top69.7 %[−46.5, 185.9] %Free
Bottom8.1 %[5.2, 10.9] %Free
IC508.18 µM[0.284, 236] µMFree
pIC505.09± 0.53 (SE)Derived
Hill slope (nH)1.16[−0.19, 2.51]Free

R² 0.988RMSE 1.3 %df 4Fit engine 1.2.0 · 8 / 8 points

Review decision

IC50

8.18 µM

95 % CI 0.284–236 µM · pIC50 5.09

Electronically signed

mlaurent (mlaurent)

2026-11-04 17:40:00 UTC

Meaning: Approved — result valid for SAR use

record sha256 2c87…4de6

This result is immutable. A correction creates a new version with a mandatory reason and its own signature.

History

append-only
  1. v1 auto-fit createdsystem · 2026-11-04 17:05:40 UTCFit engine 1.2.0 · 8 / 8 points · incompleteTop, wideCi
  2. Re-authenticatedmlaurent · 2026-11-04 17:40:00 UTCPassword verified for signing
  3. Signed — Approvedmlaurent · 2026-11-04 17:40:00 UTCMeaning: Approved — result valid for SAR use · v1 · auto-fit · record hash sha256 2c87…4de6