PL-2026-0487 · H1975 cells — Cell viability assay
QC passed96-well · ASY-H1975-01 v1Read 2026-11-04 16:42:19 UTCInstrument EnVision-02 (mock import)Raw file sha256 a6c0…e8d9Imported by mlaurent
- PL-2026-0485Z′ 0.85Passed
- PL-2026-0486Z′ 0.85Passed
- PL-2026-0487Z′ 0.84Passed
- PL-2026-0488Z′ 0.85Passed
- PL-2026-0489Z′ 0.90Passed
- PL-2026-0490Z′ 0.82Passed
6,52351,370RLU · raw
HC
Cmpd A · 4-pt, 1:3 from 10 µM
Cmpd B · 4-pt, 1:3 from 10 µM
LC
123456789101112
ABCDEFGH
Masked — excluded from QC and fits, retained
Edge effect (HC, outer rows vs interior): +3.9 %
96 wells · 0 masked
Plate QC
rules QC-STD v3Z′-factor0.84accept ≥ 0.50
- Signal window (SW)
- 17.7≥ 2.0
- Signal / background
- 6.87≥ 3.0
- CV high control
- 4.0 %≤ 10 %
- CV low control
- 4.2 %≤ 10 %
- Masked wells
- 0.0 %≤ 5 %
Plates failing any rule are rejected automatically. Rejection can be overridden only by a signed exception.
Well
Pick a well on the plate to inspect it.
Audit trail — this plate
Open full audit trail| Timestamp (UTC) | Actor | Event | Detail |
|---|---|---|---|
| 2026-11-04 16:42:19 UTC | system | QC evaluated — passed | QC-STD v3 · all rules met |
| 2026-11-04 16:42:19 UTC | system | Normalization v1 computed | μHC, μLC from cols 1–2 / 11–12 (n = 16 each) |
| 2026-11-04 16:42:19 UTC | system | Raw data imported | EnVision-02 export PL-2026-0487.csv · 96 wells · sha256 a6c0…e8d9 · stored write-once |