■ Demo— not a validated GxP system — synthetic data

Affinis

PL-2026-0501 · H1975 cells — Cell viability assay

QC passed

96-well · ASY-H1975-01 v1Read 2026-11-04 16:42:19 UTCInstrument EnVision-02 (mock import)Raw file sha256 8283…ec88Imported by mlaurent

6,51852,832RLU · raw
HC
Cmpd A · 4-pt, 1:3 from 10 µM
Cmpd B · 4-pt, 1:3 from 10 µM
LC
123456789101112
ABCDEFGH

Masked — excluded from QC and fits, retained

Edge effect (HC, outer rows vs interior): +4.8 %

96 wells · 0 masked

Plate QC

rules QC-STD v3
Z′-factor0.85accept ≥ 0.50
Signal window (SW)
19.1≥ 2.0
Signal / background
6.88≥ 3.0
CV high control
3.8 %≤ 10 %
CV low control
3.6 %≤ 10 %
Masked wells
0.0 %≤ 5 %

Plates failing any rule are rejected automatically. Rejection can be overridden only by a signed exception.

Well

Pick a well on the plate to inspect it.

Audit trail — this plate

Open full audit trail
Audit trail — this plate
Timestamp (UTC)ActorEventDetail
2026-11-04 16:42:19 UTCsystemQC evaluated — passedQC-STD v3 · all rules met
2026-11-04 16:42:19 UTCsystemNormalization v1 computedμHC, μLC from cols 1–2 / 11–12 (n = 16 each)
2026-11-04 16:42:19 UTCsystemRaw data importedEnVision-02 export PL-2026-0501.csv · 96 wells · sha256 8283…ec88 · stored write-once