PL-2026-0417 · EGFR L858R/T790M — TR-FRET kinase assay
QC passed384-well · ASY-EGFR-TM-03 v2Read 2026-11-04 16:42:19 UTCInstrument EnVision-02 (mock import)Raw file sha256 6cbb…8681Imported by mlaurent
- PL-2026-0415Z′ 0.86Passed
- PL-2026-0416Z′ 0.86Passed
- PL-2026-0417Z′ 0.86Passed
- PL-2026-0418Z′ 0.34Rejected
- PL-2026-0419Z′ 0.83Passed
- PL-2026-0420Z′ —Awaiting read
6,28549,687RFU · raw
HC
Cmpd A · 10-pt, 1:3 from 10 µM
Cmpd B · 10-pt, 1:3 from 10 µM
LC
123456789101112131415161718192021222324
ABCDEFGHIJKLMNOP
Masked — excluded from QC and fits, retained
Edge effect (HC, outer rows vs interior): +4.5 %
384 wells · 1 masked
Plate QC
rules QC-STD v3Z′-factor0.86accept ≥ 0.50
- Signal window (SW)
- 20.6≥ 2.0
- Signal / background
- 6.82≥ 3.0
- CV high control
- 3.5 %≤ 10 %
- CV low control
- 3.8 %≤ 10 %
- Masked wells
- 0.3 %≤ 5 %
Plates failing any rule are rejected automatically. Rejection can be overridden only by a signed exception.
Well
Pick a well on the plate to inspect it.
Audit trail — this plate
Open full audit trail| Timestamp (UTC) | Actor | Event | Detail |
|---|---|---|---|
| 2026-11-04 17:03:12 UTC | system | QC evaluated — passed | QC-STD v3 · all rules met |
| 2026-11-04 17:03:12 UTC | system | Normalization v2 computed | Excludes D15 · v1 retained, read-only |
| 2026-11-04 17:03:11 UTC | mlaurent | Well D15 masked | “Dispense error — Dispense error — tip dispense error flagged in instrument log for D15” |
| 2026-11-04 16:42:19 UTC | system | QC evaluated — passed | QC-STD v3 · all rules met |
| 2026-11-04 16:42:19 UTC | system | Normalization v1 computed | μHC, μLC from cols 1–2 / 23–24 (n = 32 each) |
| 2026-11-04 16:42:19 UTC | system | Raw data imported | EnVision-02 export PL-2026-0417.csv · 384 wells · sha256 6cbb…8681 · stored write-once |